Specifications
SB-1150 Installation & Operation Manual - Scale Configuration 17
7.2 Default Menu
The default menu is used to return the scale back to its factory settings and is shown in Figure 7-4.
Figure 7-4. Default Menu
Use the following steps to return the settings back to their factory default.
1. Press the ENTER key. The display shows a default value of NO.
2. To change to YES, press the Up/Down arrow keys.
3. Press the ENTER key and the display shows DONE.
4. Press the BMI key to return to PROG/CALIB/DEF/DONE/TEST upper level menu.
Selecting YES and pressing the ENTER key will reset to factory defaults settings without changing the
calibration and will return you to weigh mode.
7.3 Scale Calibration
Use the following steps to calibrate the scale.
1. Turn on scale by pressing the On/Off key and the ENTER key simultaneously.
2. The unit cycles through its startup function and continues to display the software version. Continue to
hold both keys until
Id appears.
3. Access the setup key located in the back of the scale to enter the setup parameters for the scale. You can
use a small paper clip, small screwdriver or other similar object to press the setup key.
Figure 7-5. Setup Switch Location on Back of Indicator
4. Prog appears on the display. Press the BMI key to toggle along the parameter menu.
5. Calib appears on the display and enter the calibration parameters.
6. Press the ENTER key and a numeric value is displayed which represents the amount of weight that is used
for calibration.
Lb will be flashing. To switch between lb and kg, press the Up/Down arrow keys on the
display. Once a unit is selected, press
ENTER and the right most digit will be flashing.
7. To change the calibration load value, use the Up/Down arrow keys to increment/decrement the flashing
digit.
8. Use the BMI key to move the flashing digit to the left or right.
DEFCALIBPROG
DONE
TEST
NO
YES
Note
Setup switch location
on back of indicator.