7.0 IQ9500 Counting Scale Setup
The following paragraphs contain instructions for
initial IQ9500 setup and operation either in the battery
or the AC power mode. These instructions include
setting the specication data bits for both internal and
external scale congurations.
7.1 Setting Up the IQ9500
When setting up the IQ9500 counting scale or adding
a second channel, third channel or a force balance, it
is necessary to assign a different location for each
scale whether it is present or not.
Specications 16 and 17 are used only to set up the
annunciators for each of the four scales. These
specications do not turn on the additional scales.
Scales 2 and 3 are turned on and off in the weights and
measures specications contained in Specication 25,
bit 1 (
Scale Type) and Specication 32, bit 0 (Scale
Connected to the Cable Port 2 Connector). The
force balance is turned on and off in Specication 8,
bit 3 (
RS-232C Connection, Force Balance) and the
connector type is selected in Specication 13, bits 2,
1, and 0 (
RS-232 Connector).
Table 7-1 contains the default bit congurations for
Specications 16 and 17.
NOTE: Specifications 16 and 17 can only be changed by
an authorized service technician.
Do not set Specifications 16 and 17 to the configuration shown below. A lockup condition will result if
these configurations are selected. Contact an authorized service technician if a lockout condition
occurs.
Should the IQ9500 become locked up in all 888888s
as a result of invalid specication settings, an
authorized service technician can only perform the
following procedure:
1. Unplug scale from AC outlet.
2. Press and hold the
3 and 9 keys while
plugging in AC cord to outlet.
3. The display will show
S-on. Release the 3 and
9 keys.
4. Press and hold
REZERO key while pressing
141, press + key to advance to SPEC 16.
Change SPEC 16 and SPEC 17 to assign a
different location for each scale whether it is
present or not.
Specication Default Setting Bits 2 and 3 Bits 0 and 1
16 0001 Scale 1
00: Internal Scale 1
01: Internal Scale 2
10: External scale
11: Force balance
Scale 2
00: Internal Scale 1
01: Internal Scale 2
10: External scale
11: Force balance
17 1011 Scale 3
00: Internal Scale 1
01: Internal Scale 2
10: External scale
11: Force balance
Scale 4
00: Internal Scale 1
01: Internal Scale 2
10: External scale
11: Force balance
Table 7-1. Specifications 16 and 17 Default Settings
Specication Default Setting Bits 2 and 3 Bits 0 and 1
16 0010 Scale 1
00: Internal Scale 1
Scale 2
10: External scale
17 1000 Scale 3
10: External scale
Scale 4
00: Force balance
!
Caution