6.0 Calibration
The calibration procedure maintains the scale
accuracy within specications and can serve as a
performance test procedure. Your IQ9500 scale should
be turned on for a minimum of 10 minutes and the
platform excercised three to four times before
attempting to calibrate.
NOTE: Use the SCALE key to select the platform to
calibrate. The scale number is displayed on the lower
secondary display. For dual-platform scales the
calibration procedure described below must be repeated
for each scale.
6.1 IQ9500 Display Resolution
The following tables list the weight resolution
capabilities for the IQ9500 counting scale. Table 6-1
has the single channel platform conguration while
Table 6-2 contains the dual channel scales. Use these
tables to verify specication settings before
calibration.
Capacity, lb
Weighing
Resolution
Specication 20 Specication 21 Specication 23 Specication 25 Specication 32
0.5 0.0001 0101 0100 0100 0000 1010
1 0.0001 0101 0100 0000 0000 1010
2.5 0.0002 0001 0100 0000 0000 1010
5 0.0005 1001 0100 0000 0000 1010
10 0.001 0101 0011 0000 0000 1010
25 0.002 0001 0011 0000 0000 1010
50 0.005 1001 0011 0000 0000 1010
100 0.01 0101 0010 0000 0000 1010
Table 6-1. Single Channel Platform
Capacity, lb
Weighing
Resolution
Specication
20
Specication
21
Specication
22
Specication
23
Specication
25
Specication
32
0.5/10 0.0001/0.002 0100 0100 0011 0100 0010 1010
0.5/25 0.0001/0.005 0110 0100 0011 0100 0010 1010
0.5/50 0.0001/0.01 0101 0100 0010 0100 0010 1010
1.0/10 0.0001/0.001 0101 0100 0011 0000 0010 1010
1.0/25 0.0001/0.002 0100 0100 0011 0000 0010 1010
1.0/50 0.0001/0.005 0110 0100 0011 0000 0010 1010
2.5/10 0.0002/0.001 0001 0100 0011 0000 0010 1010
2.5/25 0.0002/0.002 0000 0100 0011 0000 0010 1010
2.5/50 0.0002/0.005 0010 0100 0011 0000 0010 1010
5.0/10 0.0005/0.001 1001 0100 0011 0000 0010 1010
5.0/25 0.0005/0.002 1000 0100 0011 0000 0010 1010
5.0/50 0.0005/0.005 1010 0100 0011 0000 0010 1010
Table 6-2. Dual Channel Scales