Users Manual

CS200, Users
August, 2009
11
How to calibrate the CS200
The following details the calibration procedure for the Intercomp CS200 crane scale.
There are four parameters that can be set before entering the actual calibration. When
changing a value, advance at least to “save
savesave
save” before turning off so that your changes are
saved.
Step Description Note Recommended
rd rt
rd rtrd rt
rd rt
Read rate 1 = 1 reads/sec
2 = 2 reads/sec
3 = 4 reads/sec
4 = 7 reads/sec
5 = 10 reads/sec
6 = 18 reads/sec
7 = 25 reads/sec
4
AZT
AZTAZT
AZT
Auto Zero Tracking 0 = disabled, 1,2,3 3
dely
delydely
dely
calibration delay 0 = no delay
1 = 7 second delay for cal
points
1
grad
gradgrad
grad
graduation 4 = 5 lb
5 = 2 lb
6 = 1 lb
7 = 0.5 lb
8 = 0.2 lb
9 = 0.1 lb
10 = 0.05 lb
11 = 0.02 lb
12 = 0.01 lb
cap
capcap
cap
capacity
save
cal
calcal
cal
calibration type 1 =1 point cal linear
0101 = 3 point cal
0101
LL00
LL00LL00
LL00
Zero read
LL01
LL01LL01
LL01
Cal load 1 Enter load 1
LL02
LL02LL02
LL02
Cal load 2 Enter load 2
LL03
LL03LL03
LL03
Cal load 3 Enter load 3
Start up.
1. Turn the power OFF.
2. Remove the calibration blocking shunt to allow entry into calibration mode.
3. Turn the power ON.
4. Wait for scale to warm up (3 minutes from power on).
5. Press ZERO and BACKLIGHT together and release to enter the calibration
mode.
The scale shows “rd rt
rd rtrd rt
rd rtí
íí
í. The read rate is the number of internal reads per
second the scale performs. A higher setting will result in a faster responding