Error Messages
[Minus signs cycle across the display]: The CS1500 has entered sleep mode.
Press any key or change the weight to return to normal weighing.
NOTE: a pressed key here will perform its designated function as well as exit the scale
from sleep mode.
[Minus signs fill the display]: The CS1500 is waiting for a stable reading to continue.
LB
LBLB
LB
Low batteries. This message blinks, and if ignored too long the unit will
shut itself off.
OE
OE OE
OE
The scale is over capacity or outside the A/D converter range. Reduce the
load to the scale.
DISE
DISE DISE
DISE
Display error, the scale is unable to display the number completely. Press
zero to return the weight reading to zero.
EEPE
EEPEEEPE
EEPE
EEPROM error, the scale has had it's calibration corrupted or destroyed;
the scale will require calibration.
OL
OLOL
OL
Zero overload; the CS1500 has attempted to zero a reading outside of its
zero-range limit. This message can occur only when the “Zero range” or
“Canadian specifications” is turned on. See EE-11
EE-11EE-11
EE-11 and EE-12
EE-12EE-12
EE-12 in the
“calibration” section.
E-0
E-0E-0
E-0
This message occurs if the scale is turned on with a load applied which is
greater than +/- 10% of capacity. Return the scale's load to zero. This only
occurs when “Canadian specifications” or “Initial zero range” is turned on.
See EE-12
EE-12EE-12
EE-12 and EE-13
EE-13EE-13
EE-13 in the “calibration” section.