Users Manual User guide

CS1500, Users
Rev S, June, 2009
17
Calibration switch
The calibration of the scale is protected from accidental change by a shunt placed on
“CAL” (instead of “RUN”) of J8, located on the top/middle of the A/D 20 BIT circuit
board.
Enabling the Calibration
The shunt must be moved from “RUN” to “CAL”. When you are done calibrating,
place the shunt back on “RUN” to protect the calibration from change.
How to calibrate the scale
The following details the calibration procedure for the crane scale. There are eight
parameters that can be set without moving the calibration blocking switch, followed by
five more parameters and calibration that require the calibration blocking switch be in
the enabled position.
Step Display Parameter Note
Default
4
ee-ee
ee-eeee-ee
ee-ee
Skip
0=no skip, 1=EE-10
EE-10EE-10
EE-10, 2=LL-00
LL-00LL-00
LL-00
0
5
EE-00
EE-00EE-00
EE-00
Sample Rate 1 to 64 4
6
EE-01
EE-01EE-01
EE-01
Update Rate 1 to 32 4
7
EE-02
EE-02EE-02
EE-02
Demand Output 1=Yes, 0=Continuous 1
8
EE-03
EE-03EE-03
EE-03
Baud Rate 0 to 9 0
9
EE-04
EE-04EE-04
EE-04
Auto-off Time 0 to 255; 0=off 20
10
EE-05
EE-05EE-05
EE-05
Power up in KG 1=kg, 0=lb 0
11
EE-06
EE-06EE-06
EE-06
Tare Type 1=keyboard, 0=push-button 0
12
EE-07
EE-07EE-07
EE-07
Sleep Mode Time 0 to 255; 0=off 5
13
ee-08
ee-08ee-08
ee-08
Set Point 1 1 to 99999 99999
14
ee-09
ee-09ee-09
ee-09
Set Point 2 1 to 99999 99999
Information saved
Check for calibration blocking switch
15
EE-10
EE-10EE-10
EE-10
AZT 0=off,1=0.6,2=1,3=3 2
16
EE-11
EE-11EE-11
EE-11
Zero Range 0=off,1=on 0
17
EE-12
EE-12EE-12
EE-12
Canadian Specifications 0=off, 1=on 0
18
EE-13
EE-13EE-13
EE-13
Initial Zero Range 0=off, 1=on 0
19
EE-14
EE-14EE-14
EE-14
Graduation 0 to 11 6
20
Information saved
21
LL-00
LL-00LL-00
LL-00
Zero read Enter capacity
22
LL-01
LL-01LL-01
LL-01
First weight Enter first weight
23
LL-02
LL-02LL-02
LL-02
Second weight Enter second weight
24
LL-03
LL-03LL-03
LL-03
Third weight Enter third weight
25
Information saved
Start up
1. Move the calibration blocking switch to the “CAL” position if you intend to
calibrate.