Specifications
26 QB Series Operation Manual
1 - ERR
• The calibration weight is less than 1/2
full capacity
• Reinitiate calibration with a test mass equal
to the nominal capacity of the balance
• Consult your local Rice Lake Weighing
Systems distributor
2-ERR
• The data error exceeds 1%
• Defective weighing mechanism
• Consult your local Rice Lake Weighing
Systems distributor
3-ERR
4-ERR
• Calibration has started with a load on
the platter
• Remove load and reinitiate calibration
process
• Consult your local Rice Lake Weighing
Systems distributor
U - ERR
• Material wedged underneath the
platform support plate
• Remove platform and platform support
plate, then clean and reassemble
Condition Cause Corrective Action
Table 6-1. Troubleshooting Chart (Continued)