Specifications
Function Parameters 15
4.0 Function Parameters
The parameter menu allows you to activate or deactivate certain features and functions of the QB Series balance.
The following section provides information on the functions of the QB and how to view and change those
functions. Table 4-1 below describes the various functions of the QB.
NOTE: Bold type indicates factory setting.
Display Setting Function Description
1.SET
1 Mode Selection Weighing mode only
2 Counting mode only
2.SEL
1 Comparator
Function
Comparator OFF
2 Comparator ON
21.CO
1 Conditional
Response
Parameter
(available only when
comparator is set
to ON)
Active when balance is stable and in motion
2 Active when balance is stable ONLY (recommended)
22.LI
0 Range Response
Parameter
(available only when
comparator is set
to ON)
Inactive for data near zero and for negative data (recommended)
1 Active for the entire weighing range (all data)
23BU
0 Response
Condition (available
only when the
comparator is set
to ON)
Solid arrow next to appropriate display legend
1 Flashing arrow next to UNDER display legend.
Solid arrow next to ACCEPT and OVER display legends.
2 Flashing arrow next to ACCEPT display legend.
Solid arrow next to UNDER and OVER display legends.
3 Flashing arrow next to OVER display legend.
Solid arrow next to UNDER and ACCEPT display legends
4 Flashing arrow next to UNDER and ACCEPT display legends.
Solid arrow next to OVER display legend.
5 Flashing arrow next to ACCEPT and OVER display legends.
Solid arrow next to UNDER display legend.
6 Flashing arrow next to ACCEPT, UNDER, and OVER display legends.
3.AO
0 Auto Zero Auto zero function off
1 Auto zero function on (± 3d)
4.AP
0 Sleep Mode Sleep mode ON (available only when internal rechargeable battery installed).
Turns device off after three minutes of inactivity when running off of battery power.
1 Sleep mode OFF
5.rE
1 Response Speed Very fast (extremely stable environment)
2 Fast (stable environment)
3Average (typical environment)
4 Slow (difficult environment)
5Very slow (unstable working environment)
6.IF
0 Baud rate (available
only when optional
RS-232 is installed).
Interface off or not installed
1 Continuous serial output (6-digit)
2 Continuous serial output (7-digit)
Table 4-1. Function List