Specifications

Troubleshooting 31
11.0 Troubleshooting
The following section contains a troubleshooting chart for the IB Series scale. Refer to this chart to locate causes
and countermeasures to various conditions.
Table 11-1. Troubleshooting Chart
Condition Cause Countermeasure
Limit function data
cannot be set
Scale is not set to Limit Function mode
Upper and lower limits have not been
set, or lower limit upper limit setting
has been entered
[#] mark will flash at OVER/ACCEPT/
UNDER positions.
Check that limit function settings are
correct: See page 27
Display will not
stabilize
•Wind or vibration is affecting scale Set up scale elsewhere
Scale is set on unstable surface Set scale level
Extended usage or moving scale has
affected span value
Perform span adjustment
Displayed value is in
error. Weight value
deviating
Mis-performed tare subtraction Reset tare subtraction
Scale is not set level Set scale level
Extended usage or moving scale has
affected span value
Perform span adjustment
Linearity
error
Mechanical adjustment has changed
due to some modification in scale
characteristics, etc
Consult your local Rice Lake Weighing
Systems Service distributor
Display is blank Adapter is not connected
Battery needs recharging
Auto Power
OFF is activated
Recharge battery
•Press the
ON/OFF
Scale will not weigh
up to rated capacity
•Tare + sample weight exceeds capacity
•(Weight range = maximum capacity -
tare weight)
Recheck tare weight
The LOW-BATT lamp
is lit
If the battery option is employed, the
voltage of the internal batteries has
dropped
Recharge batteries or connect the AC
adapter
The SLEEP lamp is lit If the battery option is employed, the
displayed value has not been used for
more than three minutes
•Press a key or weigh an item
b - ERR
Electronic malfunction due to static
electricity electromagnetic interference
Consult your local Rice Lake Weighing
Systems distributor
0 - ERR
Capacity is exceeded
•Tare is overweight
(Weighing value = sample weight + tare weight)
Something contacting and holding up pan or pan base
U - ERR