User Manual
76 M-Class
Advanced Entry Calibration (continued)
Step Operator Action Displayed Message Comment
13
Us ing the
FWD
Key, select t he
Gain Nu mber det ermined in Step
12. Pr ess the
Key to
select the setting.
!
HBJO!USBO!!!!!!=zzz?+1
9!!!!!!!!!)1!.!42*!
!
Or, for reflective media:
HBJO!SFGM!!!!!!=zzz?+1
9!!!!!!!!!)1!.!42*!
S elect ion is denoted wit h an ‘*’.
14
Pl ac e the medi a in the Med i a
S ensor. Rec ord the sens or
rea ding and la bel it ‘ P’
(pa per).
P l ace the b acking, ma rk, or
not c h in the M edia Sensor.
R ecord t he s ens or reading and
label it ‘G’ or ‘M ’ (Gap or
Mar k).
R emove all media. Recor d
t he s ens or r eading and label it
‘E’ (Emp t y).
!
HBJO!USBO!!!!!!=zzz?+1
9!!!!!!!!!)1!.!42*!
!
Or, for reflective media:
HBJO!SFGM!!!!!!=zzz?+1
9!!!!!!!!!)1!.!42*!
Wher e ‘ yyy’ is a numeric al va lue rep resenti ng the
cu rrent sensor rea ding.
15
Press the
Key. T hen
press the
FWD
Ke y.
!
BEWBODFE!FOUSZ!
TFOTPS!MFWFMT!
!
The s ens or readings must be manually entered int o
the printer now.