User Manual Rev. E - Version 2.5 User guide

Configuration 31
Scale Calibration Menu
Calibration can be performed in two places within the menu: the CALIBR menu shown in Figure 3-6 on page 26 is
an in-depth scale setup and calibration.
A “quick access” calibration is shown in Figure 4-1 on page 55.
Display and edit
test weight value
Same as POINT 1
Previous A/D raw
counts are shown
Press calibrate
softkey to calibrate
zero. Press menu
to cancel
calibrating.
Previous A/D raw
counts are shown
Press calibrate
softkey to calibrate
zero. Press menu
to cancel
calibrating.
Calibrating, please wait
Press enter to
remove offset from
zero and span
calibrations
POINT 1
POINT 2 POINT 3
POINT 4
POINT 5
WZERO WVAL WSPAN REZEROWLIN
WGT 1
CAL1
CALIBR
Figure 3-9. Scale Calibration Menu Layout
Table 3-8. Scale Calibration Menu Parameters
Parameter Choices Description
WZERO Allows you to view or change the zero calibration AD count value.
WVAL Allows you to view and edit the test weight value.
WSPAN Allows you to view or change the span cal
ibration AD
count value.
WLIN POINT 1-5 Press ENTER to display and edit test weight and
calibration values for up to five linearization
points. Perform linear calibration only after WZERO and WSPAN have been set.
Parameters include CAL1 — View or change linear poi
nt
calibration value.
WGT1 — Sets test value.
REZERO Removes offset from zero and span calibrations. Can be
used during a span calibration
process using hooks and chains.