6.1.4 Clear and Reset Commands
The following commands can be used to clear and reset the Counterpart:
RS: Reset system. Resets the indicator without resetting the configuration.
RESETCONFIGURATION: Restores all configuration parameters to their default values (menu mode only). The
RESETCONFIGURATION function can also be initiated by pressing navigating to the DEFALT parameter under the
VERS menu and selecting YES. Then press
ENTER to reset the indicator.
Note
All load cell calibration settings are lost when the RESETCONFIGURATION command is run.
CLEARBUFFER: Clears the Weighvault data buffer.
6.1.5 Parameter Setting Commands
Parameter setting commands allow you to display or change the current value for a particular configuration
parameter.
Current configuration parameter settings can be displayed in all
modes using the following syntax:
command<ENTER>
Most parameter values can be changed in menu mode only.
Use the following command syntax when changing parameter
values: command=value<ENTER>, where value is either
a number or a parameter value. Use no spaces before or after the equal (=) sign. If you type an incorrect command,
the display reads
??.
For example, to set the motion band parameter to
5 divisions, type the following:
SC.MOTBAND=5<ENTER>
For parameters with selectable values, enter the command and equal sign followed by a question mark:
command=?<ENTER> to see a list of those values. The indicator must be in menu mode to use this function.
Note
Some parameters are valid only if other parameters or parameter values are specified. Restrictions for front-
panel configuration also apply to serial command configuration.
Command Description Values
SC.GRADS#n Graduations 1–100000
SC.ZTRKBND#n Zero track band 0, 0–100
SC.ZRANGE#n Zero range 1.900000, 0–100
SC.MOTBAND#n Motion band 1, 0–100
SC.SSTIME#n Standstill time 1–65535
SC.OVRLOAD#n Overload FS+2%, FS+1D, FS+9D, FS
SC.DIGFLTR1#n
SC.DIGFLTR2#n
SC.DIGFLTR3#n
Digital filtering 1, 2, 4, 8, 16, 32, 64, 128, 256
SC.DFSENS#n Digital filter cutout sensitivity 2OUT, 4OUT, 8OUT, 16OUT, 32OUT,
64OUT
, 128OUT
SC.DFTHRH#n Digital filter cutout threshold NONE, 2D, 5D, 10D, 20D, 50D, 100D,
200D, 2
50D
SC.THRESH#n Zero threshold 0999
9999
SC.SMPRAT#n Sample rate 30HZ, 60HZ, 120HZ, 240HZ, 480HZ,
960H
Z
SC.PWRUPMD#n Power up mode GO, DELAY
SC.TAREFN#n Tare function BOTH, NOTARE, PBTARE, KEYED
SC.PRI.DECPNT#n Primary units decimal position 8.888888, 88.88888, 888.8888, 8888.888,
8888
8.88, 888888.8, 8888888, 8888880,
8888800
Table 6-4. Scales Serial Commands