User Manual Rev. D - Version 2.4 Instruction Manual

26 Counterpart Counting Scale Indicator
3.6 Calibration Menu
See Section 4.0 on page 55 for calibration procedures. The Calibration menu can be protected by assigning a
password in the Feature menu.
Note
The Counterpart requires the WZERO and WSPAN points to be calibrated. The linearity points are optional,
but must NOT duplicate zero or span. During calibration, the T
ARE key acts as a data entry confirmation key. It
also acts as an EXECUTE key and accepts the value if calibration was successful.
Display and edit
test weight value
Same as POINT 1
Previous A/D raw
counts are shown
Press calibrate
softkey to calibrate
zero. Press menu
to cancel
calibrating.
Previous A/D raw
counts are shown
Press calibrate
softkey to calibrate
zero. Press menu
to cancel
calibrating.
Calibrating, please wait
Press enter to
remove offset from
zero and span
calibrations
POINT 1
POINT 2 POINT 3
POINT 4
POINT 5
WZERO WVAL WSPAN REZEROWLIN
WGT 1
CAL1
ID DISPLAYTESTSETUPAUDIT CALIBR SETPTS
Figure 3-6. Calibration Menu Layout
Table 3-5. Calibration Menu Parameters
CALIBR Menu
Parameter Choices Description
Level 2 submenus
WZERO Press the calibrate softkey to display
the AD raw counts.
Press calibrate softkey again to calibrate zero, or press MENU to cancel. “CALIBRATING” will
a
ppear prior to automatically moving to WVAL.
WVAL Press the calibrate softkey to display and edit the test weight
value. Press the calibrate softkey
again to move to WSPAN.
WSPAN Press the calibrate softkey to display the AD raw counts. Pr
ess the calibrate softkey to
calibrate the span or press MENU to cancel. “CALIBRATING” will appear prior to automatically
moving to WLIN.
WLIN POINT 1 —
POI
NT 5
WGT 1 allows you to display and edit the test weight value. Press the calibrate softkey to edit
the value. CAL1 allows you to calibrate and display the raw AD value. Press TARE to perform
calibration. Press the calibrate softkey to move to the next calibration point.
Note
The linearity points are optional, but must not duplicate zero or span. They
must be between zero and span.
REZERO Press the calibrate softkey to remove an offset val
ue from the zero and span calibrations if
hooks or chains are being used during calibration.
Note
Use this parameter only after WZERO and WSPAN have been set. See
Section 4.1 on page 55 for more information.