User Guide
Table Of Contents
Tools required: T10 screwdriver X1 , tweezer X1
Notice:
Pls turn off the device,disconnect the power adapter before maintenance
Operation Procedure :
→Remove the defective sieve bed
→Disconnect the valve base from sieve bed column
→Connect the valve base to the new sieve bed
→Install the new sieve bed to concentrator
Stage 1 Standby
Remove the battery by forcing down the latch ,then slide out the battery horizontally.
Remove the anti-slip mats installed on
the screw holes on the bottom of the
concentrator.
Remove the two fixing screws on the
bottom of base by using the T10
screwdriver.