Owner's Manual
Operation Instructions
22
6.
Set to the flow rate prescribed by your physician or
clinician.
Press the + or - setting buttons to adjust the P2-E6 to the desired flow rate. The
current setting can be viewed on the display from 1 to 6.
CAUTION
Make sure the power is in a well-ventilated place. During operation, the power supply may
get hot. Make sure the power supply is cool before handling.
7.
Wear the nasal cannula on your face and breathe through your nose.
The P2-E6 will sense if you are breathing from it. If you are not
yet breathing through the cannula, the P2-E6 will begin to pulse
automatically about every 3 seconds.
As soon as you begin breathing through the cannula, the device
will begin delivering pulses based on your breathing. As your
breathing rate changes, the P2-E6 will sense these changes
and adjust the amount of oxygen at your next inhale.
WARNING
If you feel discomfort using the device, consult your doctor immediately.
CAUTION
A Low O2:
!
< 87% alert will display on the screen if the oxygen level
drops below
recommended levels. If the alarm persists, contact your provider.
CAUTION
The display screen will dim if there is no operation for 30 seconds. You can press any
button to light up the display.
CAUTION
The P2-E6 will notify you with an audible alert (if enabled) and a display showing "Absence
Of Breath" if no breath has been detected for 15 seconds. After 15 seconds, the device will
enter auto-pulse mode until breath is detected. Once breath is detected, the device will
resume normal delivery of oxygen.
General Operation-continued