Owner's Manual

15
Alerts - continued
Alarm item
Alarm condition
System process
Display of screen
Valve Check Fail
Valve switch failed
shut down after
10s
Valve Check Fail:
Pls Contact Provider
Cooling Fan Fail
Cooling fan failed
shut down after
10s
Cooling Fan Fail:
Pls Contact Provider
Battery Depleted
RSOC: ! 5%
Without adapter
shut down after
10s
Battery Depleted:
Replace battery or
connect to adapter
System Cold
System temperature <
32°F/0°C
shut down after
10s
System Cold:
Pls Shut down, Move
to warmer place
Battery Cold
Battery
temperature <
32°F/0°C
shut down after
10s
Battery Cold:
Pls Shut down, Move
to warmer place
System Hot
System
temperature >
140°F/60°C
shut down after
10s
System Hot:
Pls Shut down, Move
to cooler place
Battery Hot
Battery
temperature >
14F/65°C
shut down after
10s
Battery Hot:
Pls Shut down,
Only use adapter
Gas Supply Fail
Flow or
concentration below
normal after injection
shut down after
10s
Gas Supply Fail:
Pls Contact Provider
System Startup
Fail
Concentration less
than 87%
continuously > 15S
after system startup
shut down after
10s
System Startup Fail:
Pls Contact Provider
Power Supply
Fail
System voltage
< 10 5v
shut down after
10s
Power Supply Fail:
Pls Contact Provider