5. MAINTENANCE, ADJUSTMENTS AND CALIBRATION
Page 83 ©2001~2007 DH Instruments, a Fluke Company
5.2.2.5 <4VALVE>
PURPOSE
To view and adjust the device channel valve voltage signal.
See Sections 5.2.2.1 and 5.2.1 for information on MFC-CB device channel
signal calibration.
OPERATION
To view the device channel valve voltage signal and the current valve voltage
adder (A) and multiplier (M), press [SPECIAL], <4cal>, select the desired device
channel, press <4valve>, <1view>. The display is:
1. Valve voltage value with the Valve adder and
multiplier applied (V).
2. Indication that this is a display of the Valve
signal.
3. Device channel indicator.
4. Current value of the multiplier applied to the
Valve reading.
5. Current value of the adder (V or mA) applied
to the Valve reading.
While in the view screen, the [SET MFC] key is active so that set point values
can be changed.
To adjust the value of the valve voltage adder (A) and/or multiplier (M), press
[ESCAPE] to return to the view/cal screen, and then press <2cal> to access a
screen in which the values of A and M can be edited. After entering the edited
values, press [ESCAPE] and return to <1view> to view the valve voltage values
with the edited calibration coefficients applied.
See Section 5.2.1 for an explanation of adders and multipliers and their use in
adjusting electrical settings and readings.
5.2.2.6 <5CAL DATE>
PURPOSE
To view and edit the calibration date for the MFC-CB device channel.
OPERATION
Press [SPECIAL] and select <4cal>. After selecting the device channel,
select <5date>. The date format is YYYYMMDD.
There is only one calibration date for both device channels (DEV1 and DEV2).
1.7443 V Valve
A
0.00000V M1.0000 D1
1
2
3
45