©2001~2007 DH Instruments, a Fluke Company Page 80
5.2.2.2 <1SENSE>
PURPOSE
To view and adjust the readings of the device channel sense line.
See Sections 5.2.2.1 and 5.2.1 for information on MFC-CB device channel
signal calibration.
OPERATION
To view the reading of the sense line and the adder(A) and multiplier(M), press
[SPECIAL], <4cal>, select the desired device channel, and press
<1sense>, <1view>. The display is:
1. Reading of the device channel Sense with the
Sense cal adder and multiplier applied.
2. Indication that this is a display of the Sense signal.
3. Device channel indicator.
4. Current value of the multiplier applied to the
Sense reading.
5. Current value of the adder (always in V)
applied to the Sense reading.
While in the view screen, the [SET MFC] key is active so that set point values
can be changed (see Section 3.2.8).
To adjust th
e value of the adder (A) and/or multiplier (M) press [SPECIAL],
select the desired device channel, and press <1sense>, <2cal> to access a
screen in which the values of A and M can be edited. After entering the edited
values, press [ESCAPE] and return to <1sense>, <1view> to view the sense
line reading with the edited calibration coefficients applied.
See Section 5.2.1 for an explanation of adders and multipliers and their use in
adjusting electrical settings and readings.
5.2.2.3 <2MEAS> (MEASURE)
PURPOSE
To view and adjust the device channel measure signal.
See Sections 5.2.2.1 and 5.2.1 for information on MFC-CB device channel
signal calibration.