Specifications

4. REMOTE OPERATION
Page 73 ©2001~2007 DH Instruments, a Fluke Company
VSENSE(n)
Purpose
Read the voltage sensed at the MFC/MFM terminal.
Syntax “VSENSEn
Arguments n:
Device number (optional). Uses the active device if not
given.
'1' to address DEV1
'2' to address DEV2
Remarks
The MFC-CB uses a voltage sense line to measure the voltage present at
an external MFC. This sense voltage is used to re-adjust the voltage sent to
the MFC to compensate for line losses.
Example Command:
Reply:
“VSENSE1
“2.00037 V"
Errors ERR# 10:
ERR# 43:
The prefix "n" is invalid
Not in volts mode
See Also 3.2.8, 7.1.1, "MFCCH", "VOUT"
VVALTEST(n)
Purpose
Read voltage at MFC valve test terminal.
Syntax “VVALTESTn
Arguments n:
Device number (optional). Uses the active device if not
given.
'1' to address DEV1
'2' to address DEV2
Remarks
The MFC-CB provides an input to measure the MFC valve test voltage.
This measurement is referenced to the MFC –15 Volt supply.
Example Command:
Reply:
“VVALTEST1
“6.2371 V"
Errors ERR# 10: The prefix "n" is invalid
See Also "MFCCH"