Specifications
4. REMOTE OPERATION
Page 71 ©2001~2007 DH Instruments, a Fluke Company
TEXTn=
Purpose Specify text to be written to the MFC-CB's front panel display.
Syntax “TEXTn=text”
Arguments n:
text:
The position on the front display to display the text (1 to 40)
The text to display
Remarks
The MFC-CB normally uses the same displays in remote mode as in local
operation, but the display can also be used to indicate user text. The
"TEXTn=" command clears the display if the normal measurement data is
still being displayed before displaying the text. Subsequent use of this
command simply overwrites the current user text.
Reverting to local operation or selecting a display type ("DISP=" command)
will return the display to normal operation.
Example Command:
Reply:
“TEXT21=The bottom line" (displays on the bottom line)
“The bottom line""
Errors ERR# 2:
ERR# 10:
The data given is too large
The prefix "n" is invalid
See Also "TEXT:CLRn=", "DISP="
TIME(=)
Purpose Read or set the internal clock time.
Syntax “TIME”
“TIME=hh:mmxx
Arguments hh:mmxx:
The time to set in the format where hh is the hours from 1 to
12, mm is the minutes from 1 to 59 followed by “am” or “pm”.
Example
Comman
Reply:
“TIME=1:22am”
“01:22am”
Errors ERR# 7: The time given is invalid
See Also 3.4.3.3
VER
Purpose Read the MFC-CB embedded software version.
Syntax “VER”
Remarks
The embedded software version of the MFC-CB can be read. This is useful
for checking for the presence of the MFC-CB and for reference purposes.
Example Command:
Reply:
“VER”
“DH INSTRUMENTS, INC MFC-CB Ver1.00”