Specifications
MFC-CB™ Operation and Maintenance Manual
©2001~2007 DH Instruments, a Fluke Company Page 70
SCRSAV(=)
Purpose Read or set the front panel screen saver period.
Syntax “SCRSAV=period”
“SCRSAV”
Default “SCRSAV=10”
Arguments Period: The period of inactivity before the display dims.
Remarks The MFC-CB display will dim after a period of inactivity.
Example Command:
Reply:
“SCRSAV=60”
“60 ”
Errors ERR# 6: The data given is invalid
See Also 3.4.3.1
SN
Purpose Read the serial number of the MFC-CB.
Syntax “SN”
Remarks
Each MFC-CB is serialized. This serial number is also imprinted on the
product label on the bottom of the MFC-CB and is displayed in the power up
introductory screen.
Example Command:
Reply:
“SN”
“SN”
TEXT:CLRn=
Purpose Clear text from the front panel.
Syntax “TEXT:CLRn=size”
Arguments n:
size:
The position on the front display to start clearing (1 to 40)
The number of positions to clear (41 - n)
Remarks
The MFC-CB normally displays the measurement data while in remote, but
can also be used to indicate user text. The "TEXT:CLRn=" command is
used to clear sections of user text from the screen.
Reverting to local operation or selecting a display type ("DISP=" command)
will return the display to normal operation.
Example Command:
Reply:
“TEXT:CLR1=20
“20"
Errors ERR# 6:
ERR# 10:
The data given is invalid
The prefix "n" is invalid
See Also "TEXTn=", "DISP"