Specifications

4. REMOTE OPERATION
Page 69 ©2001~2007 DH Instruments, a Fluke Company
POWER=
Purpose Turns the MFC-CB's soft power ON/OFF.
Syntax “POWER=state”
Arguments state: ‘0’ to power the MFC-CB OFF
‘1’ to power the MFC-CB ON
Remarks
The MFC-CB will automatically power up when it receives a remote
command while in soft power OFF.
Example Command:
Reply:
“POWER=1”
“POWER=1”
Errors ERR# 7: The state argument is not a '0' or a '1'
See Also 3.1.2.3
REMOTE
Purpose Enable remote local lockout operation.
Syntax “REMOTE”
Remarks
A “REMOTE” command deactivates the front panel. All front panel controls
are disabled. The “REMOTE” command can only be canceled by a
“LOCAL” command or by turning the MFC-CB power OFF and then ON.
Example Command:
Reply:
“REMOTE”
“REMOTE”
See Also “LOCAL”
RESET
Purpose Reset the MFC-CB to the default operating parameters.
Syntax “RESET”
Remarks
The “RESET” command can be given to return certain MFC-CB settings to a
default state. This reset corresponds to the RESET - SETS function from
the front panel.
Example Command:
Reply:
“RESET”
“RESET”
See Also 3.4.1.1, "*RST"
RES(n)(=)
Purpose Read or set the front panel display resolution.
Syntax “RESn=%FS”
“RESn
Arguments
n:
%FS:
Device channel number (optional). Uses the active device
channel if not given.
'1' to address DEV1
'2' to address DEV2
The %FS resolution (1 to 0.01 %).
Defaults “RESn=0.010%”
Remarks
The display resolution affects only the front panel display resolution.
Remote responses always have full resolution.
Example Command:
Reply:
“RES1=0.1”
“0.100”
Errors ERR# 6:
ERR# 10:
The data given is invalid
The prefix "n" is invalid
See Also 3.3.3