Specifications

MFC-CB™ Operation and Maintenance Manual
©2001~2007 DH Instruments, a Fluke Company Page 62
ADJ:MEAS(n)(=)
Purpose Read or set the user measurement adjustment setting.
Syntax “ADJ:MEASn
“ADJ:MEASn=adder, mult”
Arguments
n:
adder:
mult:
Device channel number (optional). Uses the active device
channel if not given.
'1' to address DEV1 settings
'2' to address DEV2 settings
The measurement adjustment adder in %FS (0 to 100 %)
The adjustment multiplier
Defaults “ADJ:MEASn=0.0000%, 1.0000”
Remarks
The user measurement adjustment affects the measurement signal and is
used to compensate for line loss or other imperfections in the MFC/MFM or
MFC/MFM interface when measuring MFC/MFC voltage or current. The
measurement adjustment is device channel specific. Do not confuse this
adjustment with the MFC-CB calibration coefficients.
Example Command:
Reply:
“ADJ:MEAS2=3, 1.003”
“0.0003%, 1.0030”
Errors ERR# 6:
ERR# 10:
The data given is invalid
The prefix "n" is invalid
See Also 3.3.5, “ADJ:SET”, "DEV"
ADJ:SET(n)(=)
Purpose Read or set the user set adjustment setting.
Syntax “ADJ:SETn
“ADJ:SETn=adder, mult”
Arguments
n:
adder:
mult:
Device channel number (optional). Uses the active device
channel if not given.
'1' to address DEV1 settings
'2' to address DEV2 settings
The set adjustment adder in %FS (0 to 100 %)
The set adjustment multiplier
Defaults “ADJ:SETn=0.0000%, 1.0000”
Remarks
The user set adjustment affects the set point signal and is used to
compensate for line loss or other imperfections in the MFC interface when
setting the MFC voltage or current. The measurement adjustment is device
channel specific. Do not confuse this adjustment with the MFC-CB
calibration coefficients.
Example Command:
Reply:
“ADJ:SET2=3, 1.003”
“0.0003%, 1.0030”
Errors ERR# 6:
ERR# 10:
The data given is invalid
The prefix "n" is invalid
See Also 3.3.5, "DEV", "ADJ:MEAS"