Specifications

4. REMOTE OPERATION
Page 61 ©2001~2007 DH Instruments, a Fluke Company
ACAL:SET(n)(=)
Purpose Read or set the set point calibration coefficients.
Syntax “ACAL:SETn
“ACAL:SETn=Vadder, Vmult, Cadder, Cmult”
Arguments
n:
Vadder:
Vmult:
Cadder:
Cmult:
Device channel number (optional). Uses the active device
channel if not given.
'1' to address DEV1 settings
'2' to address DEV2 settings
The voltage set adder in Volts
The voltage set multiplier
The current set adder in mA
The current set multiplier
Defaults “ADJ:SETn=0.0000 V,1.00000, 0.0000 mA,1.00000”
Remarks
The set calibration coefficients affect the set point signal output. There are
separate coefficients for the voltage and current outputs and for each device
channel.
For the Voltage calibration, note that the sense calibration ("ACAL:SENSE"
command) better determines the voltage actually set.
Example Command:
Reply:
“ACAL:SET2=0.01, 1, 0.05, 1”
“0.0100 V,1.00000, 0.0500 mA,1.00000”
Errors ERR# 6:
ERR# 10:
The data given is invalid
The prefix "n" is invalid
See Also 5.2.2.1, 5.2.2.4, "DEV", "ACAL:SENSE", "ACAL:MEAS", "ACAL:DATE"
ACAL:VALVE(n)(=)
Purpose Read or set the valve calibration coefficients.
Syntax “ACAL:VALVEn
“ACAL:VALVEn=adder, mult”
Arguments
n:
adder:
mult:
Device channel number (optional). Uses the active device
channel if not given.
'1' to address DEV1 settings
'2' to address DEV2 settings
The valve voltage adder in Volts
The valve voltage multiplier
Defaults “ADJ:VALVEn=0.0000 V,1.00000”
Remarks The valve calibration coefficients affect the valve voltage measurement.
Example Command:
Reply:
“ACAL:VALVE2=0.01, 1”
“0.0100 V,1.00000”
Errors ERR# 6:
ERR# 10:
The data given is invalid
The prefix "n" is invalid
See Also 5.2.2.1, 5.2.2.5, "DEV", "ACAL:SET", "ACAL:DATE"