Specifications
MFC-CB™ Operation and Maintenance Manual
©2001~2007 DH Instruments, a Fluke Company Page 60
ACAL:MEAS(n)(=)
Purpose Read or set the measurement calibration coefficients.
Syntax “ACAL:MEASn”
“ACAL:MEASn=Vadder, Vmult, Cadder, Cmult”
Arguments
n:
Vadder:
Vmult:
Cadder:
Cmult:
Device channel number (optional). Uses the active device
channel if not given.
'1' to address DEV1 settings
'2' to address DEV2 settings
The voltage measurement adder in Volts
The voltage measurement multiplier
The current measurement adder in mA
The current measurement multiplier
Defaults “ADJ:MEASn=0.0000 V,1.00000, 0.0000 mA,1.00000”
Remarks
The measurement calibration coefficients affect the measurement signal.
There are separate calibrations for voltage and current and for active device
channel 1 and 2.
Example Command:
Reply:
“ACAL:MEAS2=0.01, 1, 0.05, 1”
“0.0100 V,1.00000, 0.0500 mA,1.00000”
Errors ERR# 6:
ERR# 10:
The data given is invalid
The prefix "n" is invalid
See Also 5.2.2.1, 5.2.2.3, "DEV", "ACAL:SET", "ACAL:DATE"
ACAL:SENSE(n)(=)
Purpose Read or set the sense calibration coefficients.
Syntax “ACAL:SENSEn”
“ACAL:SENSEn=adder, mult”
Arguments
n:
adder:
mult:
Device channel number (optional). Uses the active device
channel if not given.
'1' to address DEV1 settings
'2' to address DEV2 settings
The sense voltage adder in Volts
The sense voltage multiplier
Defaults “ADJ:SENSEn=0.0000 V,1.00000”
Remarks
The sense calibration coefficients affect the voltage sense measurement.
This in turn affects the voltage set operation as the sense measurement is
used to set the voltage output as close as possible to the set point value,
compensating for line loss.
Example Command:
Reply:
“ACAL:SENSE2=0.01, 1”
“0.0100 V,1.00000”
Errors ERR# 6:
ERR# 10:
The data given is invalid
The prefix "n" is invalid
See Also 5.2.2.1, 5.2.2.2, 7.1.1, "DEV", "ACAL:SET", "ACAL:DATE"