Specifications
3. OPERATION
Page 37 ©2001~2007 DH Instruments, a Fluke Company
• Adjustment adder to 0 and multiplier to 1 (see Section 3.3.5).
• MFC Switchbox to none (see Section 3.4.6).
•
Screen saver activation to 10 minutes (see Section 3.4.3.1).
•
Keypad sounds to medium (see Section 3.4.3.2).
3.4.1.2 <2UNITS>
PURPOSE/OPERATION
To access Reset - Units, press [SPECIAL], <1reset>, <2units>.
Reset - Units clears and sets to default all UNIT OF MEASURE functions.
These include:
• Six flow units of measure selectable in MFC profile editing to defaults
(see Section 3.3.4).
•
DEV1 and DEV2 display unit to electrical signal (see Section 3.2.2).
3.4.1.3 <3MFC>
PURPOSE/OPERATION
Use special caution with this reset as it will delete up to 97 custom MFC profile
entries which will have to be reentered if deleted inadvertently.
To access Reset - MFC, press [SPECIAL], <1reset>, <3MFC>.
Reset - MFC clears and sets to default all MFC profile functions. These include:
• DEV1 and DEV2 to MFC profile #1 (see Section 3.2.6).
•
Clear all custom MFC profiles (see Section 3.3.1).
3.4.1.4 <4CAL>
PURPOSE/OPERATION
Use special caution with this reset as critical calibration data may be altered.
To access Reset - Cal, press [SPECIAL], <1reset>, <4cal>.
Reset - Cal clears and sets to default the user calibration values (see
Section 5) including:
•
DEV1 and DEV2 sense, measure, set and valve:
Adder: 0
Multiplier: 1
• DEV1 and DEV2 calibration date to 19980101.