Specifications

MFC-CB™ Operation and Maintenance Manual
©2001~2007 DH Instruments, a Fluke Company Page 20
The cursor is on the active DISPLAY function. Selecting a DISPLAY function returns to
the run screen with the selected DISPLAY function active.
See PRINCIPLE above and Sections 3.2.4.1 through 3.2.4.6 for details on each
DISPLAY func
tion.
The default DISPLAY function is CLEAN which causes the second line of the display to be
empty except for the active device channel indicator at the far right (see Section 3.1.1).
3.2.4.1 <1DEV1&2>
PURPOSE
To view both MFC-CB device channels simultaneously.
See Section 3.2.4, PRINCIPLE.
OPERATION
To activate the DEV1&2 DISPLAY, press [DISPLAY] and select <1DEV1&2>.
Selecting <1DEV1&2> returns to the MAIN run screen with the DEV1&2
DISPLAY active.
With the DEV1&2 DISPLAY active, the MAIN run screen is:
1. Active device channel indicator.
2. Standard set point and measurement line for
DEV1.
3. Standard set point and measurement line for
DEV2.
DEV1 is always on the top line and DEV2 is always on the bottom line,
regardless of which device channel was active when the DEV1&2 DISPLAY
was selected. The device channel indicator <
Æ> indicates which device channel
is active for set point changes and other functions. Press [DEV1/2] to toggle the
active device channel indicator between DEV1 and DEV2.
To go to a DISPLAY other than DEV1&2, press [DISPLAY] and make a new
DISPLAY choice.
Æ
4.0000 V 3.9985
2.0000 V 2.0036