Specifications
MFC-CB™ Operation and Maintenance Manual
©2001~2007 DH Instruments, a Fluke Company Page 92
The following configuration can be used with:
• Brooks model 5850E, and 5851E
Table 15. Brooks MFC Connector Pin Out
MFC-CB DEV1, DEV2
25 PIN DSUB
CONNECTOR PIN NO.
SIGNAL
DESCRIPTION
MFC CARD
EDGE
CONNECTOR
PIN NO.
MFC 15 PIN DSUB
CONNECTOR PIN
NO.
1
Case Ground 1 (A1) 14
2
Supply Common C (B3) 9
3
MFC Measure (+) 3 (A3) 2
4
+15 Volts 4 (A4) 5
7
Set Point Sense (-) 2 (A2) 10
11
Valve Test D (Optional) (B4) 7 (Optional)
14
MFC Set point (+) A (B1) 8
15
MFC Set point (-) 2 (A2) 10
16
MFC Measure (-) B (B2) 10
17
Valve Test Common F (Optional) (B6) 10 (Optional)
19
-15 Volts F (B6) 6
20
Set Point Sense (+) A (B1) 8
7.3 DRIVERS
The MFC-CB drivers option provides eight open collector drivers for operating external valves,
solenoids, indicators, etc. When operating from the setup-driver screen (see Section 3.2.7), pressing
[ENTE
R] allows the operating mode of the drivers to be set. The two modes of operation are
Momentary and Toggle. A momentary driver changes state while the corresponding driver number
on the keyboard is being pressed. In toggle mode, the driver state toggles each time the
corresponding key is pressed and released.
Each output can sink 500 mA at 12 V. However, the total output of all the activated drivers cannot
exceed 1 Amp. Therefore, if multiple drivers are being activated, refer to Table 16 as a guide.
Table 16. Dri
ver Output
# OF ACTIVE
DRIVERS
MAX CURRENT
PER OUTPUT
1 500 mA
2 400 mA
3 275 mA
4 200 mA
5 160 mA
6 135 mA
7 120 mA
8 100 mA