Specifications
MFC-CB™ Operation and Maintenance Manual
©2001~2007 DH Instruments, a Fluke Company Page 2
1.2.2 ANALOG OUTPUT SPECIFICATIONS
VOLTAGE OUTPUT
Range
0 to 6 VDC
Uncertainty
± 0.015 %FS
Resolution
0.1 mVDC
Output Impedance (max)
100 Ω per channel
Load Impedance (min)
10 kΩ per channel
Short Circuit Protection
Yes (indefinite)
CURRENT OUTPUT
Range
4 to 20 mA
Uncertainty
± 0.025 %FS
Resolution
0.4 μA
Maximum Load
500 Ω (compliance voltage: 10 VDC)
Short Circuit Protection
Yes (indefinite)
1.2.3 ANALOG INPUT SPECIFICATIONS
VOLTAGE INPUT (MEASURE AND SENSE)
Range
0 to 6 VDC
Min to Max Measurable
-0.25 to 6.00 VDC
Uncertainty
± 0.015 %FS
Resolution
0.1 mVDC
Input Impedance (min)
1 MΩ
CURRENT INPUT (MEASURE)
Range
4 to 20 mA
Min to Max Measurable
4 to 20 mA
Uncertainty
± 0.025 %FS
Resolution
0.4 μA
Input Impedance
250 Ω
1.2.4 VALVE TEST POINT SPECIFICATIONS
Range
+2 to +15 VDC (in reference to -15 VDC)
Accuracy
± 0.1 %FS
Resolution
0.3 mVDC