Specifications
molbox™ RFM™ OPERATION AND MAINTENANCE MANUAL
© 1998-2007 DH Instruments, a Fluke Company Page 82
Table 17. Security Levels - Functions NOT Executed Per Function/Level
(Continued)
KEYS LOW MEDIUM HIGH
[SPECIAL], <5prefs>, <3time> (make changes)
• • •
[SPECIAL], <5prefs>, <4ID>, <2edit>
• • •
[SPECIAL], <5prefs>, <5log> (clear log)
• • •
[SPECIAL], <6remote>
•
[SPECIAL], <6remote> (changes settings)
• •
[SPECIAL], <7micro>
• •
[SPECIAL], <8head>
• •
[SPECIAL], <9BPR> (change mode)
• •
Remote communications disabled
•
OPERATION
molbox RFM is delivered with NO active password so access to the User
Level menu is open. The user level is set to <1Low>. User levels can be
changed freely until a password has been created. RESET functions (see
Section
3.6.1) do not affect the password setting.
To access the USER LEVEL function, press [SPECIAL], <2level>.
If NO password yet exists or if the
correct password has been entered,
the display is:
1change user level
3edit password
Selecting <1change user level> brings
up the restriction menu:
Restriction: 1none
2low 3medium 4high
Select the desired restriction level, or press [ESCAPE] to return to the current
run screen.
Selecting <2edit password> displays the
user password and allows it to be edited.
Passwords can be up to six numbers in
length and cannot start with a zero.
Password: pppppp
0 disables password
If 0 is entered as the password value, then the password is made inactive and a
password will NOT be required to access the user level menu. This is the factory
default with a security level of <2low>.
Once a password has been entered, the user level cannot be changed without
reentering the password.