Specifications

molbox™ RFM™ OPERATION AND MAINTENANCE MANUAL
© 1998-2007 DH Instruments, a Fluke Company Page 78
3.6.1.1 <1SETS>
PURPOSE/OPERATION
To access Reset - Sets, press [SPECIAL] and select <1reset>, <1sets>.
Reset - Sets clears and sets to default the user settings for various measurements.
These include:
Flow unit of measure to sccm (see Section
3.4.3).
Pressure unit of measure to kPa (see Section
3.5.2).
Temperature unit of measure to °C (see Section 3.5.3).
Gas type to N2 (see Section 3.4.2).
Stability criterion to 0.5 sccm (see Section 3.5.5).
K factor to 1 (see Section 3.4.1).
DISPLAY function to Rate (see Section
3.4.6.1).
RPT Tare value to 0 (see Section 3.4.4.1).
Tare preference to upstream pressure (see Section 3.4.4.1.1).
Flow adder to 0 and flow mu
ltiplier to 1 (see Section 3.5.6).
Resolution to 0. 01 % (see Section 3.4.9).
BPR mode to Auto (see Section 3.6.9)
3.6.1.2 <2UNITS>
PURPOSE/OPERATION
To access Reset - Units, press [SPECIAL] and select <1reset>, <2units>.
Reset - Units clears and sets to default all UNIT OF MEASURE functions. These include:
Six flow units of measure selectable from [UNIT] to defaults (see Section
3.4.3).
Flow unit
of measure to sccm (see Section 3.4.3).
R
eference temperature for uxxx units of measure to 0 °C (see Section 3.4.3.3).
Volume flow unit of measure conditions
to molbloc for temperature and
standard atmospheric pressure for pressure (see Section 3.4.3.4).
Pressure unit of measure to kPa (see Section 3.5.2).
User pressure unit coefficient to 1.00/Pa (see Section 3.5.2).
Temperature unit of measure to °C (see Section 3.5.3).