Specifications
molbox™ RFM™ OPERATION AND MAINTENANCE MANUAL
© 1998-2007 DH Instruments, a Fluke Company Page 78
3.6.1.1 <1SETS>
PURPOSE/OPERATION
To access Reset - Sets, press [SPECIAL] and select <1reset>, <1sets>.
Reset - Sets clears and sets to default the user settings for various measurements.
These include:
• Flow unit of measure to sccm (see Section
3.4.3).
• Pressure unit of measure to kPa (see Section
3.5.2).
• Temperature unit of measure to °C (see Section 3.5.3).
• Gas type to N2 (see Section 3.4.2).
•
Stability criterion to 0.5 sccm (see Section 3.5.5).
• K factor to 1 (see Section 3.4.1).
• DISPLAY function to Rate (see Section
3.4.6.1).
• RPT Tare value to 0 (see Section 3.4.4.1).
• Tare preference to upstream pressure (see Section 3.4.4.1.1).
• Flow adder to 0 and flow mu
ltiplier to 1 (see Section 3.5.6).
• Resolution to 0. 01 % (see Section 3.4.9).
•
BPR mode to Auto (see Section 3.6.9)
3.6.1.2 <2UNITS>
PURPOSE/OPERATION
To access Reset - Units, press [SPECIAL] and select <1reset>, <2units>.
Reset - Units clears and sets to default all UNIT OF MEASURE functions. These include:
• Six flow units of measure selectable from [UNIT] to defaults (see Section
3.4.3).
• Flow unit
of measure to sccm (see Section 3.4.3).
• R
eference temperature for uxxx units of measure to 0 °C (see Section 3.4.3.3).
• Volume flow unit of measure conditions
to molbloc for temperature and
standard atmospheric pressure for pressure (see Section 3.4.3.4).
• Pressure unit of measure to kPa (see Section 3.5.2).
• User pressure unit coefficient to 1.00/Pa (see Section 3.5.2).
• Temperature unit of measure to °C (see Section 3.5.3).