Specifications
molbox™ RFM™ OPERATION AND MAINTENANCE MANUAL
© 1998-2007 DH Instruments, a Fluke Company Page 68
With the FREEZE DISPLAY active, the MAIN run screen is:
1. Standard MAIN run screen top line.
2. Flow measured in the current flow units when
[ENTER] was pressed (displays 0.00 by
default when FREEZE DISPLAY is first
activated).
* 101.75 sccm N20
F 99.24
Pressing [ENTER] causes the current flow measured by molbox RFM to be
captured and displayed.
If the flow measurement unit is changed while the FREEZE DISPLAY is active,
the FREEZE value defaults back to zero.
To go to a DISPLAY other than FREEZE, press [DISPLAY] and make a new
DISPLAY choice.
3.4.6.8 <8CLEAN>
PURPOSE
To activate the CLEAN DISPLAY.
See Section 3.4.6, PRINCIPLE.
OPERATION
To activate the CLEAN DISPLAY press [DISPLAY] and select <8clean>.
Selecting <8clean> returns to the MAIN run screen with the CLEAN DISPLAY active.
With the CLEAN DISPLAY active, the MAIN run screen is:
1. Standard MAIN run screen top line.
2. “Clean” second line.
* 101.45 sccm N2O
To go to a DISPLAY other than CLEAN, press [DISPLAY] and make a new
DISPLAY choice.
3.4.7 [MICRO] (OPTIONAL)
PURPOSE
To turn ON and OFF automatic operation of the optional MICRORANGE FLOW
MEASUREMENT function.