Specifications

molbox™ RFM™ OPERATION AND MAINTENANCE MANUAL
© 1998-2007 DH Instruments, a Fluke Company Page 64
3.4.6.3 <3 HI/LO>
PURPOSE
To activate the HI/LO DISPLAY.
See Section 3.4.6, PRINCIPLE.
OPERATION
To activate the HI/LO DISPLAY press [DISPLAY] and select <3hi/lo>. Selecting
<3hi/lo> resets the HI/LO values and returns to the MAIN run screen with the
HI/LO DISPLAY active. With the HI/LO DISPLAY active, the MAIN run screen is:
1. Standard MAIN run screen top line.
2. Highest flow observed since HI/LO reset.
3. Lowest flow observed since HI/LO reset.
* 101.22 sccm N20
H 101.44 L99.113
The HI/LO values change each time a new HI or LO flow value occurs.
The HI/LO record can be reset at any time by pressing [ENTER] allowing a
HI/LO reset without going back through the DISPLAY menu.
Changing the flow unit of measure, the gas, the K factor or running a TARE
function while in HI/LO resets the HI/LO record.
To go to a DISPLAY other than HI/LO, press [DISPLAY] and make a new
DISPLAY choice.
3.4.6.4 <4TOTAL> (TOTALIZER)
PURPOSE
To activate the TOTALIZER DISPLAY.
See Section 3.4.6, PRINCIPLE.
OPERATION
To activate the TOTALIZER DISPLAY, press [DISPLAY] and select <4total>.
The display is:
1. Edit field for time over which to totalize
(hh:mm:ss). Default period is 00:10:00;
maximum 99:59:59.
Totalizing period:
00:10:00
Edit the totalizing period as desired. Pressing [ENTER] returns to the MAIN run
screen with the TOTALIZER DISPLAY active. With the TOTALIZER DISPLAY
active the MAIN run screen is: