Specifications

3. OPERATION
Page 55 © 1998-2007 DH Instruments, a Fluke Company
RPTs is available. Though it may not be practical and generally is not
necessary, the best possible reference with which to run AutoZ in absolute
measurement mode is a gas operated piston gauge (such as a DHI PG7601)
applying an absolute pressure near atmospheric pressure to the molbox
RFM test port. The best day to day reference is a properly calibrated DHI
RPM4 with a BA100K RPT interfaced directly as an external device to the
molbox RFM COM2 port.
Allow the molbox RFM to stabilize at atmospheric pressure and ambient
temperature for 10 to 15 minutes before running AutoZ.
If AutoZ is on, the AutoZ value will be applied while running the calibration of
molbox RPTs and an AutoZ indication is included in the run calibration screen
(See Section 5.2.4.1).
OPERATION
To access the molbox RFM AutoZ function press [TARE], <4AutoZ>. The
display is:
1. Active RPT designator.
2. Indication of whether AutoZ is currently ON or
OFF for this RPT and measurement mode.
1off 2view
3edit 4run ON
Select <1off> (or <1on>) to change the AutoZ status.
Select <2view> to view the current values of P
offset
for the two RPTs.
P
offset
should be zero when the molbox RFM is new or has just been calibrated.
Select <3edit> to edit the values of Poffset.
The value of P
offset
is always displayed and entered in Pascal (Pa).
Select <4run> to run the AutoZ routine which determines and activates P
offset
values by measurement of P
std,0
(see Section 3.4.4.4.2).
3.4.4.4.1 Edit AutoZ
The edit AutoZ function should be used with great caution as entering
inappropriate values and turning ON AutoZ may result in out of tolerance
measurements. In normal operation, the value of the AutoZ offset, P
offset
,
should be changed using the run AutoZ function (see Section 3.4.4.4.2).
Before editing P
offset
, (see Section 3.4.4.4, PRINCIPLE).