Specifications

molbox™ RFM™ OPERATION AND MAINTENANCE MANUAL
© 1998-2007 DH Instruments, a Fluke Company Page 146
5.6.1.1 UPSTREAM ABSOLUTE RPT
Measures absolute pressure on the UP (HI) molbloc pressure connection port.
Connected pneumatically to the valving assembly and electrically to the main
board (COM4) (see Section 5.6.1.8).
5.6.1.2 DOWNSTREAM ABSOLUTE RPT
Measures absolute pressure on the DOWN (LO) molbloc pressure connection
port. Connected pneumatically to the valving assembly and electrically to the
main board (COM5) (see Section 5.6.1.8)
5.6.1.3 DIFFERENTIAL MICRORANGE RPT (OPTIONAL)
Measures differential pressure between the UP (HI) and DOWN (LO) molbloc
pressure connection ports (upstream (+), downstream (-)). Connected
pneumatically (x2) to the valving assembly and electrically to the main board
(COM6) (see Section 5.6.1.8).
5.6.1.4 DISPLAY
2 x 20 vacuum fluorescent, alpha-numeric display mounted to front panel.
5.6.1.5 POWER SUPPLY
molbox RFM has one power supply with two voltage outputs:
V DC (± 5 %) @ 0.9 Amps: for valve excitation, RPT supply and analog circuit.
V DC (± 5 %) @ 2.5 Amps: for micro board, main board, display and fan.
5.6.1.6 MICRO BOARD
The micro board supports a Motorola 68302 micro-controller, EPROM,
EEPROM, 128k x 16 bit NVRAM, 8 Mbit flash memory; RS232 and IEEE-488.2
communications; keypad and display control. An I/O port controls other ports
and devices within molbox RFM.
5.6.1.7 VALVING MODULE
The valving module (see Figure 17) consists of a manifold equipped with three
12 V, 2-way, solenoid valves. The valves connect and disconnect the RPTs to
the UP (HI) and DOWN (LO) rear panel ports and bypass the two ports together.
If molbox RFM is equipped with the microrange option, a fourth (3-way) solenoid
valve puts the microrange differential RPT into and out of service (bypassed or
not bypassed).