Specifications
4. REMOTE OPERATION
Page 121 © 1998-2007 DH Instruments, a Fluke Company
REMOTE
Purpose Enable remote local lockout operation.
Syntax “REMOTE”
Remarks A REMOTE command deactivates the front panel. All front panel
controls will be disabled. The REMOTE command can only be
canceled by a LOCAL command or by turning OFF the molbox
RFM power then re-applying it.
Example Command:
Reply:
“REMOTE”
“REMOTE”
See Also “LOCAL”
RES(=)
Purpose Read or set the local flow display resolution.
Command
Query
“RES=n”
“RES”
Default “RES=0.01%”
Arguments n: Resolution in %FS (1% - 0.01%)
Remarks The molbox’s front panel resolution can be specified with this
command. Displayed resolution becomes 10 times smaller than the
current RES setting (ie. default = 0.001% FS) when the microrange
option is active. The resolution of remote flow measurement replies
are always 0.001% FS (0.0001% FS with microrange).
Example
Sent:
Query reply:
“RES=.01”
“0.010%”
Errors ERR# 6: The argument was out of bounds.
See Also 3.4.9
RESET
Purpose Reset the molbox RFM to the default operating parameters.
Syntax “RESET”
Remarks The RESET command can be given to return certain molbox
RFM settings to a default state. This reset corresponds to the
RESET - SETS function from the front panel. The reset process
can take up to 10 seconds to complete, so the user must wait this
time interval before attempting to communicate to the molbox
RFM again.
Example Command:
Reply:
“RESET”
“RESET”
See Also 3.4.1.7
SN
Purpose Read the serial number of the molbox RFM.
Syntax “SN”
Remarks Each molbox RFM is serialized. This serial number is also
imprinted on the product label on the bottom of the molbox RFM
and is displayed in the power up introductory screen.
Example Command:
Reply:
“SN”
“SN”
See Also 3.4.5.4