Specifications

molbox™ RFM™ OPERATION AND MAINTENANCE MANUAL
© 1998-2007 DH Instruments, a Fluke Company Page 114
Errors ERR# 6:
ERR# 15:
ERR# 23:
ERR# 46:
ERR# 47:
ERR# 48:
ERR# 49:
ERR# 50:
ERR# 51:
One of the arguments is not invalid.
A leak check cycle has not been started.
The microrange option is not installed in this
RFM.
A large leak exists.
The temperature changed too much to test for a
leak.
The absolute RPT disagreement too great to test
for a leak.
Possible upstream leak.
Possible downstream leak.
Possible bypass leak.
See Also 3.2.4.3, “ABORT”
LOCAL
Purpose Enable the front panel controls if in remote and go to local mode.
Syntax “LOCAL”
Remarks In LOCAL mode, all front panel operations are available. The
LOCAL command deactivates REMOTE mode.
Example Command:
Reply:
“LOCAL”
“LOCAL”
See Also “REMOTE”
MEM
Purpose Read the memory test status.
Syntax “MEM”
Remarks On power up, a memory test is run to check the integrity of the
internal data NVRAM. If the memory has been corrupted,
"FATAL MEMORY FAULT" will be displayed on power up to alert
the user, and the memory test status command will return a ‘0’.
The command will return a ‘1’ if the memory is OK.
Example Command:
Reply:
“MEM”
“0”
MICRORNG(=)
Purpose Read or set the optional microrange operation mode.
Syntax “MICRORNG=mode”
“MICRORNG”
Defaults “MICRORNG=AUTO”
Arguments mode:
“OFF”: AUTO and MANUAL MICRORANGE
functions are turned OFF.
“AUTO”: AUTO MICRORANGE function is turned
ON.
“ON”: MANUAL MICRORANGE function is turned
ON.
Remarks The AUTOMATIC and MANUAL MICRORANGE functions control
the way that the microrange option is used in the molbox RFM.
The command in it’s query form can also be used to tell if the
microrange option is installed in a molbox RFM.