Specifications
4. REMOTE OPERATION
Page 105 © 1998-2007 DH Instruments, a Fluke Company
ABORT
Purpose To stop an active averaging, tare, purge or leak check cycle.
Syntax “ABORT”
Remarks This command should be used to ensure that molbox RFM is in
a known state. The tare, purge, and leak check cycles put the
RFM into a condition where it is not usable for normal flow
measurement, so the “ABORT” command should be used
before general operation of the unit begins if the previous state
of the unit is not known. The “ABORT” command also clears out
any previous leak or tare cycle errors.
Example Command:
Reply:
“ABORT”
“ABORT”
See Also “FA”, “TARESETUP”, “LEAKCK”, “PURGE”
AIRW(=)
Purpose Read or set the humidity ratio for the gas, AIR.
Syntax “AIRW=HumidityRatio”
“AIRW”
Default “AIRW=0.00
00”
Arguments HumidityRati
o:
The humidity ratio (0-0.1)
Remarks The humidity ratio is only used if the gas is set to specify AIR. For all
other gas types, the ratio is ignored.
Example Command:
Reply:
“AIRW=.01”
“0.0100”
Errors ERR# 6: The humidity ratio is invalid.
See Also 3.2.2, 3.4.2.2
“GAS”