Specifications

molbox™ RFM™ OPERATION AND MAINTENANCE MANUAL
© 1998-2007 DH Instruments, a Fluke Company Page 102
ESR?
Purpose Read the standard event register.
Syntax
ESR?”
Query
Reply
n (0 to 255)
Remark The standard event register contents are cleared after reading.
The reply is in decimal numeric form.
Example Command:
Reply:
“*ESR?”
4”
IDN?
Purpose Identify the molbox RFM version, range, and serial number.
Syntax
IDN?”
Remarks The identification reply is made up of the manufacturer, the
model, the serial number and the software version. Each is
separated by a comma.
Query
Reply
The version string
Example Command:
Reply:
“*IDN?”
“DH INSTRUMENTS INC, molbox RFM, 1001,
Ver1.10 -dhf”
OPC(?)
Purpose Sets the operation complete bit when all operations have
completed.
Syntax
OPC”
OPC?”
Remarks This command enables the molbox RFM to set the OPC bit in
the standard event status register when it has completed all
pending functions. The query replies with a “1” when all
functions are complete.
Since the molbox RFM does not support overlapping
commands, this command has no practical use.
Query
Reply
“0” or “1”
Example Command:
Reply:
“*OPC”
“OK” (using COM1, no reply if IEEE-488 port)
Command:
Reply:
“*OPC?”
“1”