Owner`s manual
Operator manual REX-***R(F)
LISTEM Corporation 35/90
3.4 APR operation
An examination using an APR technique could consist of the following:
1. Make sure that the X- ray tube to be used is properly warmed-up.
2. Position the patient for the examination
3. Select the "Patient Size" corresponding to the patients anatomy. This operation
starts the APR mode. Select the "Pediatric" button if the patient is not an adult.
4. Select a general "Body Region" and an "Anatomical View" of the indicated on the APR
Display.
5. Technique parameters, Workstation information, Focal Spot, etc.. corresponding to the APR
selection are displayed and indicated on the control console. If needed, the parameters and
selections can be directly modified by the operator.
6. Continue with the normal procedure for a typical RAD examination.