User Manual

Confirmation
7
Pos: 3.2 /00020 BD A Software/2 0005 PMGC Ka pitelsammlung/Ü berschriften/1. 1 Überschrifte n/11 Funktions weise @ 0\ mod_12088704 93457_9.doc @ 406 @ @ 1
Method of Oper ation
2.1 Method of Operation
Pos: 3.3 /00020 BD A Software/2 0005 PMGC Ka pitelsammlung/ Warnhinweise/Hin weis nicht in PM200 verwend bar @ 0\m od_121005671 6286_9.doc @ 7 66 @ @ 1
The wireless grinding jar cover cannot be used in the PM 200.
Pos: 3.4 /00020 BD A Software/2 0005 PMGC Ka pitelsammlung/ Warnhinweise/M odul Beacht en Sie zusätzlich di e Bedienungs- Anleitung I hrer Kugelmühle! @ 0\mod_12 14301374239 _9.doc @ 172 0 @ @ 1
Comply with the operating instructions for your planetary ball mill too.
This document only has instructions for the use of the PM GrindControl. Safe
handling of the planetary ball mill is a precondition for the use of the PM
GrindControl.
Pos: 3.5 /00020 BD A Software/2 0005 PMGC Ka pitelsammlung/ 0004 Vorwort/M odul Arbeitsw eise @ 0\mo d_120885860 6206_9.doc @ 34 5 @ @ 1
This wireless grinding jar cover and the PM GrindControl software serve to
continuously acquire the parameters pressure and temperature in a grinding jar.
The measured values read by the sensors located in the grinding jar cover are
transmitted wirelessly to a receiver system in the PC.
The temperature sensor is thermally separate from the grinding jar cover and
measures only the gas temperature in the grinding chamber. As a result, even
rapid or slight changes in temperature are registered.
The transmitted data is displayed and stored in the software. For every
measurement, the values measured are stored in a file (dat-file) and are then
available for extensive evaluations.
The wireless grinding jar cover permits the use of the system in the PM 100 and
PM400.
Essentially, the following steps are necessary for using the PM GrindControl
software to monitor the temperature and the pressure in the grinding jar:
1. Definition of a measurement task
2. Integration of the measurement systems involved
3. Execution of the measurement
4. Evaluation of the measurement data
The evaluation can be conducted at a separate time from the measurement, i.e.
the raw data can be inputted again later and evaluated. The raw data contains all
the settings and protocol inputs stored in the measurement task.
Pos: 4.1 /00020 BD A Software/2 0005 PMGC Ka pitelsammlung/- - - - Seitenu mbruch - - - - @ 0\mod_12088 57688413_0.doc @ 337 @ @ 1