Manual
Cleaning, wear and service
57
Fig.
38
: Removing the pipe
5. After stopping the device you can remove and clean the pipe (TR).
6. Also clean the base (TB).
7. Reinsert the pipe (TR).
– Ensure the seal is firmly in place.
8. Screw in the sealing screw (DS).
8.4.1.2 Rinsing the cooling system
1. Fill with fresh, clean, lime-free tap water (maximum cooling agent volume
600ml).
Fig.
39
: Maximum and minimum fill levels
2. Start the device up for 1 minute at a speed of 300 revolutions per minute
(with two grinding jars).
min.
max.
Maximum cooling agent
volume: 600ml
TR
TB