User Manual
Operating the machine
40
• Set the parameters for grinding as described in the following chapters:
- Number of grinding cycles
- Precooling
- Grinding time
- Intermediate cooling time
- Grinding frequency
– It is possible to cancel at any time using the PROG button (B15)
• Press the SET button (B15).
– Pressing the SET BUTTON (B15) again stores all parameters in the
program.
5.7.2.4 Delete program
• Press the PROG button (B15) until the program to be deleted is displayed in
the memory display.
• Press the shortcut SET (B15) and - (B8) for at least 5 seconds.
Fig. 34: Deleting a program
– The program is deleted and the parameters are assigned with the values “0”
or “-”.
– It is not possible to start a deleted program.
Pos: 6.35 /00005 Ü berschrifte n/1.1 Überschrift en/1.1 Überschri ften BDA/11 Ma hl-Zyklen ei nstellen @ 0\ mod_122718525 3126_9.doc @ 38 79 @ @ 1
5.8 Setting grinding cycles
Pos: 6.36 /00010 Bedienungsa nleitungen Kapitels ammlungen/Cr yoMill II/0007 B edienung/0017 Cr yoMill II Mod ul Mahlzyklen ei nstellen @ 5\mod_13438 29253169_9.doc @ 33323 @ @ 1
Fig. 35: Setting the grinding cycles
• Switch the CryoMill on at the main switch.
– After switching on, the last used grinding parameters are displayed and can
be used.
– You can set 0 ("-") to 9 grinding cycles
NOTICE
If 0 ("-") CryoCycles are set, the grinding is conducted without nitrogen. In this
case the buttons B10, B11 and B12 are blocked.
B8