Operating the machine
36
Fig. 27: Inserting the grinding jar adapter
Pos: 6.25 /00005 Ü berschriften/ 1.1.1 Überschrift en/111 Adapter für 4/2 x 2 ml Eppendorf Rea ktionsgefäße eins etzen @ 1\ mod_1243499 921893_9.doc @ 10328 @ @ 1
5.5.4 Insert the adapter for 6 / 4 / 2 x 2-ml Eppendorf reaction vial
Pos: 6.26 /00010 Bedienungsa nleitungen Kapitels ammlungen/Cr yoMill/0101 W arnhinweis e/CryoMill HIN WEIS Zerbrechen von Eppendor f Reaktionsgefäß en @ 1\mo d_124349997578 4_9.doc @ 10 336 @ @ 1
NOTICE
Breaking of Eppendorf reaction vials
Changing material properties
– The Eppendorf reaction vials that are available as accessories change
their material properties during the cryogen grinding.
• Do not reuse reaction vials after the cryogen grinding.
• Do not insert reaction vials in centrifuges after the cryogen grinding
• Grinding may take place with a maximum 25Hz.
Pos: 6.27 /00010 Bedienungsa nleitungen Kapitels ammlungen/Cr yoMill II/0007 B edienung/0008 Cr yoMill II Mod ul Eppendorf G efäße eins etzen @ 5\mod_ 13438292403 70_9.doc @ 3 3224 @ @ 1
You can operate the CryoMill with 2, 4 or 6 Eppendorf reaction vials (2ml).
NOTICE
Grinding with Eppendorf reaction vials may only take place with a maximum 25Hz.
(frequency 1/s)
• Insert the 2ml Eppendorf reaction vials into the adapter for reaction vials as
shown in the diagram below.
Fig. 28: Inserting 2ml Eppendorf reaction vials