User Manual

Operating the machine
36
Fig. 27: Inserting the grinding jar adapter
Pos: 6.25 /00005 Ü berschriften/ 1.1.1 Überschrift en/111 Adapter für 4/2 x 2 ml Eppendorf Rea ktionsgefäße eins etzen @ 1\ mod_1243499 921893_9.doc @ 10328 @ @ 1
5.5.4 Insert the adapter for 6 / 4 / 2 x 2-ml Eppendorf reaction vial
Pos: 6.26 /00010 Bedienungsa nleitungen Kapitels ammlungen/Cr yoMill/0101 W arnhinweis e/CryoMill HIN WEIS Zerbrechen von Eppendor f Reaktionsgefäß en @ 1\mo d_124349997578 4_9.doc @ 10 336 @ @ 1
NOTICE
Breaking of Eppendorf reaction vials
Changing material properties
The Eppendorf reaction vials that are available as accessories change
their material properties during the cryogen grinding.
Do not reuse reaction vials after the cryogen grinding.
Do not insert reaction vials in centrifuges after the cryogen grinding
Grinding may take place with a maximum 25Hz.
Pos: 6.27 /00010 Bedienungsa nleitungen Kapitels ammlungen/Cr yoMill II/0007 B edienung/0008 Cr yoMill II Mod ul Eppendorf G efäße eins etzen @ 5\mod_ 13438292403 70_9.doc @ 3 3224 @ @ 1
You can operate the CryoMill with 2, 4 or 6 Eppendorf reaction vials (2ml).
NOTICE
Grinding with Eppendorf reaction vials may only take place with a maximum 25Hz.
(frequency 1/s)
Insert the 2ml Eppendorf reaction vials into the adapter for reaction vials as
shown in the diagram below.
Fig. 28: Inserting 2ml Eppendorf reaction vials