User Manual
Transport, scope of delivery, installation
14
NOTICE
Atmospheric humidity
– Electronic and mechanical components may be damaged and the
performance data alter to an unknown extent.
• Do not exceed the admissible range for atmospheric humidity.
Pos: 5.15 /00005 Ü berschriften/ 1.1 Überschrifte n/1.1 Übersc hriften BDA/11 A ufstellen des Ger ätes @ 0\ mod_12264988 49756_9.doc @ 3464 @ 2 @ 1
4.5 Installation of the machine
Pos: 5.16 /00003 Standard Kapitel/ General Modul Aufstellungsh öhe @ 0\m od_1228918538 349_9.doc @ 4 724 @ @ 1
Installation height: maximum 2000 m above sea level
Pos: 5.17 /00004 Warnhinweis e/H0055 HINW EIS CryoMill Da mpf-Fahne @ 4\mod_13140 82098654_9.doc @ 25428 @ @ 1
NOTICE
Property damage
A very cold vapour plume escapes from the nitrogen gas outlet.
– The vapour plume can greatly cool objects. This is why the humidity in the
air condenses.
• The area above the nitrogen gas outlet (B) up to the ceiling must be
kept free.
Pos: 5.18 /00004 Warnhinweis e/H0004 HINW EIS Boden Aufst ellen Vibration en Boden @ 0 \mod_1228918 882236_9.doc @ 4774 @ @ 1
NOTE
Installation
– Depending on the operating status of the mill, there may be slight
vibrations.
• Place the mill on an even, flat and balanced supporting surface only.
The supporting surface must be stable and must not vibrate.
Pos: 5.19 /00004 Warnhinweis e/H0002 HINW EIS Aufstellung Zug ang Geräteste ckdose @ 0\ mod_12338 36960983_9.doc @ 5820 @ @ 1
NOTICE
Installation of the machine
– It must be possible to disconnet the machine from the mains at any time.
• Install the machine such that the connection for the mains cable is
easily accessible.
Pos: 5.20 /00005 Ü berschriften/ 1.1 Überschrifte n/1.1 Überschrift en BDA/11 Tra nsportsicherung e ntfernen @ 0\mod_12275 30921618_9.d oc @ 3980 @ 2 @ 1
4.6 Removing the transport safeguard
Pos: 5.21 /00004 Warnhinweis e/H0018 HINW EIS Transportsich erung @ 2\ mod_127805397 2635_9.doc @ 2 0540 @ @ 1
NOTICE
Transport safeguard
– Components may be damaged.
• Operate the machine only without the transport safeguard or
transport the machine only with transport safeguard.
Pos: 5.22 /00010 Bedienungsa nleitungen Kapitels ammlungen/Cr yoMill II/000 6 Transport, Li eferumfang un d Aufstellen /Cr yoMill II Modul A uspacken und Tr ansportsicher ung entfernen @ 5\mod_1 343829053087_9. doc @ 33048 @ @ 1
• Place the device on a stable table.
• Pull the device carefully beyond the edge of the table in order to be able to
reach the screw (TA).