Manual

Operating the machine
21
Close the waste gas / overflow valve (V3).
Pos: 6.18 /00005 Ü berschriften/ 1.1 Überschrifte n/1.1 Überschrift en BDA/11 Druc kaufbau @ 1 \mod_12404 00926229_9.doc @ 8240 @ @ 1
4.7 Pressure build-up
Pos: 6.19 /00010 Bedienungsa nleitungen Kapitels ammlungen/Cr yoMill/0016 LN2 Dewar 5 0l/1604 Modul De war 50l Drucka ufbau @ 2\ mod_1267701477 789_9.doc @ 18841 @ @ 1
Close the withdrawal valve (V1) and the waste gas / overflow valve (V3).
Open the pressure build-up valve (V2).
Watch the pressure indicator (DM).
Close the pressure build-up valve (V2) as soon as the needle reaches the
red marking or nitrogen comes out of the pressure relief valve (PSV).
Fig. 6: Pressure gauge
Pos: 6.20 /00005 Ü berschriften/ 1.1 Überschrifte n/1.1 Überschrif ten BDA/11 E ntnahme von Flüs sigstickstoff @ 2\mod_126 7774957314_9. doc @ 18877 @ @ 1
4.8 Withdrawal of liquid nitrogen
Pos: 6.21 /00010 Bedienungsa nleitungen Kapitels ammlungen/Cr yoMill/0016 LN 2 Dewar 50l/16 05 Modul Dewar 50l Entna hme Flüssigstic kstoff @ 2\ mod_126770907251 7_9.doc @ 18 855 @ @ 1
Attach the liquid nitrogen line to the CryoMill and to the connection (SA).
Open the filling and withdrawal valve (V1).
After taking the liquid, close the filling and withdrawal valve (V1) again.
Pos: 6.22 /00005 Ü berschriften/ 1.1 Überschrifte n/1.1 Überschrift en BDA/11 Druc kentlastung @ 2\mod_12 67777113490 _9.doc @ 18891 @ @ 1
4.9 Pressure relief
Pos: 6.23 /00010 Bedienungsa nleitungen Kapitels ammlungen/Cr yoMill/0016 LN2 Dewar 5 0l/1606 Modul De war 50l Drucke ntlastung @ 2\ mod_1267709 470983_9.doc @ 18862 @ @ 1
Close the pressure build-up valve (V2).
Open the waste gas valve (V3).
Pos: 6.24 /00005 Ü berschriften/ 1.1 Überschrifte n/1.1 Überschrift en BDA/11 Tr ansport @ 0\m od_12264951 64391_9.doc @ 3398 @ 2 @ 1
4.10 Transport
Pos: 6.25 /00004 Warnhinweis e/V0019 VORSICHT Überdruc k im Behälter - G efärdung durc h LN2 @ 2\m od_1267775099 915_9.doc @ 1 8884 @ @ 1
CAUTION
Risks of liquid or gaseous nitrogen
Positive pressure in the vessel
The introduction of energy during transport can cause a sudden surge in
pressure. Nitrogen is blown off then and this can cause a lack of oxygen,
cryogenic burns, oxygen enrichment and associated risks.
Make sure the vessel is depressurised before you transport it,
Remove the siphon and put in the transport plugs.
Transport the vessel only in an upright position.
Avoid impacts and vibrations.
Secure the vessel against falling down, rolling away and damage.
Pos: 6.26 /00005 Ü berschriften/ 1. Überschriften/ 1 Prüfanweisu ngen @ 1\ mod_12452308964 41_9.doc @ 11 114 @ @ 1
Pos: 6.27 /00010 Bedienungsa nleitungen Kapitels ammlungen/Cr yoMill/0016 LN 2 Dewar 50l/16 12 Modul Dewar 50l Verboten ist die An wendung folgender Mi ttel und @ 2\mod_1268 231540473_9.doc @ 18957 @ @ 1
It is impermissible to use the following agents and tools:
pliers
striking tools
DM