User's Manual
iBreeze CPAP System User Manual
32
Safety Specifications
Class II
Type BF
Ingress Protection: IP22
Air filter
Air filter: Filter Efficiency: >75% (7 micron dust)
Pressure
Setting Range: 4-20 cmH
2
O
Max Single Fault Steady Pressure: 40 cmH
2
O
Static Pressure Accuracy: ±0.5 cmH
2
O (Expanded uncertainty:1.87%)
Dynamic Pressure Accuracy: ±1cmH
2
O (Expanded uncertainty:2.55%)
Pressure monitoring Accuracy: ± (2%*full scale reading + 4%*actual reading)
Flow
Test pressure (cmH
2
O)
4 8 12 16 20
19mm
breathing
tubing
Measured pressure at the
patient connection port
(cmH
2
O)
2.9 7.1 11.0 15.1 19.1
Average flow at the patient
connection port (L/min)
154.3 146.4 135.7 125.3 114.1
15mm
breathing
tubing
Measured pressure at the
patient connection port
(cmH
2
O)
2.9 6.9 10.8 14.8 18.7
Average flow at the patient
connection port (L/min)
141.4 132.7 124.2 114.3 85.4
Humidifier
Water capacity: 290 ml (MAX Water Level)
Humidity: >10 mg/L BTPS (Within the set pressure range)
Wrist Pulse Oximeter:
Model: MD300W314B4
SpO
2
Range: 70%~100%
SpO
2
Accuracy: 70%~100%: ±2%; <70%: unspecified
Pulse Rate Range: 30~250 bpm
Pulse Rate Accuracy: 30~99 bpm: ±2 bpm; 100~250 bpm: ±2%
Note: refer the user manual (version: Ver1.0w314B4) for more details.