User's Manual

iBreeze CPAP System User Manual
27
Problem Possible Cause Action
mask seems too low
(it feels like I am not
getting enough air).
progress. Ramp Time off.
I have stopped
therapy, but the
device is still blowing
air.
Device is cooling
down.
Device blows a small amount of air in
order to avoid condensation in the air
tubing. It will stop automatically after a
few minutes.
The device’s display
is erratic (crash blank
or blue screen).
The device has
been dropped or
mishandled.
Unplug the device. Reapply power to
the device. If the problem continues,
contact your home care provider.
My water tank is
leaking.
Water tank may
not be assembled
correctly.
Water tank may
be damaged or
cracked.
Check for damage and reassemble
the water tank correctly. Contact your
care provider for a replacement.
Key exception
(non-responsive or
insensitive).
Program crashes
or key
misalignment.
Unplug the device. Reapply power to
the device. If the problem continues,
contact your home care provider.
The knob is
insensitive.
Encoder is
damaged
Unplug the device. Reapply power to
the device. If the problem continues,
contact your home care provider.
The touchscreen is
not working.
The touchscreen
is damaged
Unplug the device. Reapply power to
the device. If the problem continues,
contact your home care provider.