User's Manual
iBreeze CPAP System User Manual
25
Alert Type Possible Cause Action
threshold value. parameter setting.
Low Respiratory
Rate (RR).
Alert 0 None.
Check the respiration
tube or adjust therapy
parameter setting.
Apnea. Alert 1
An apnea is detected
and exceeds the
setting duration.
Check the respiration
tube or adjust therapy
parameter setting.
High Leakage
Volume.
Alert 1
1. Inappropriate
connection of mask
and respiration tube.
2. The water tank is not
plugged in.
Check the connection
of the mask or
respiration tube, and
the connection of the
water tank.
High Inspiration
Pressure.
Alert 0
During breath cycle
monitored pressure is
higher than the set
pressure threshold.
Check the respiration
tube or adjust therapy
parameter setting.
Low Expiratory
Pressure.
Alert 0
During breath cycle
monitored pressure is
lower than the set
pressure threshold.
Check the respiration
tube or adjust therapy
parameter setting.
Mask vent holes
blocked.
Alert 1
Mask vent holes is
blocked.
Check the mask.
Low Tidal Volume
(Vt).
Alert 0
During breath cycle,
Average Vt less than
0.6*Setting Value.
Check the respiration
tube or adjust therapy
parameter setting.
System Drying Notification
The device generates
a small flow to dry the
system when it is the
standby mode after 30
minutes usage with
humidification.
No action.
Low battery
charge, reports
may not be saved.
Notification
Low button battery
power.
Please contact the
supplier if you want
reports be saved.
RTC clock failure.
Reports may not be
saved.
Alert1 RTC clock failure.
Please contact the
supplier if you want
reports be saved.
Note: Some alert messages depend on the device model.