User's Manual

iBreeze CPAP System User Manual
24
Alert Type Possible Cause Action
Non-original
card,continue using
please format.
Notification
Insert other device' SD
card in non-service
mode.
Format the SD card in
your computer and
then insert it to
device.
Software update
underway, Do not
cut off the power!
Notification Software update. No action.
System error code:
XXXX Please try to
restart, please
contact the supplier
if repeat.
Alert 2
1. Pressure sensor
failure, flow sensor
failure, blower failure in
therapy state; 2. Power
board short-circuit 3.
Humidifier heating
circuit short-circuit
Please try to restart,
please contact the
supplier if repeat.
The respiration
tube expired,
please replace.
Notification
In standby state, the
usage time exceed the
respiration tube setting
service life.
Click “Confirm”,
replace the respiration
tube.
The water tank
expired, please
replace.
Notification
In standby state, the
usage time exceed the
water tank setting
service life.
Click “Confirm”,
replace the water
tank.
The filter expired,
please replace.
Notification
In standby state, the
usage time exceed the
filter setting service
life.
Click “Confirm”,
replace the filter
The mask expired,
please replace.
Notification
In standby state, the
usage time exceed the
mask setting service
life.
Click “Confirm”,
replace the mask
It is time for device
maintenance,
Please contact the
service provider for
device
maintenance.
Notification
In standby state, the
usage time exceed the
setting device
maintenance time.
Click “Confirm”,
contact the service
provider for device
maintenance.
Low Minute
Volume (MV).
Alert 0
Minute volume less
than the setting
threshold value.
Check the respiration
tube or adjust therapy
parameter setting.
High Respiratory
Rate (RR).
Alert 0
Respiratory Rate
exceed the setting
Check the respiration
tube or adjust therapy